The CMA 11 Microdialysis Probe possesses greater spatial resolution and causes less tissue damage due to its reduced size. The probe has a cuprophane membrane with an outside diameter of 0.24 mm. The outer steel shaft diameter is 0.38 mm. The inner cannula is constructed of fused silica coated with polyimid. The inlet/outlet capillaries are mounted in a white plastic body matched to the size of a corresponding guide cannula. Once implanted, the probe sits tightly in the guide cannula´s capsule without the need for screwing or cementing. FEP tubing is connected to the probe using tubing adaptors.
|
Technical information for CMA 11
|
|
|
Membrane lenght Cuprophane
|
1, 2, 3 and 4 mm
|
|
|
Membrane diameter
|
0.24 mm
|
|
|
Stainless steel shaft diameter
|
0.38 mm
|
|
|
Shaft length
|
14 mm
|
|
|
Inlet internal volume
|
negligible
|
|
|
Outlet internal volume
|
1 µL
|
|
|
|
|
|
Ordering information for CMA 11
|
|
3/pkg
Membrane: Cuprophane
cut-off 6000 Daltons
Ref. No.
830 9581 CMA 11 Microdialysis Probe with membrane length - 1 mm
830 9582 CMA 11 Microdialysis Probe with membrane length - 2 mm
830 9583 CMA 11 Microdialysis Probe with membrane length - 3 mm
830 9584 CMA 11 Microdialysis Probe with membrane length - 4 mm
|